ProcedureCouch/Table (Longitudinal)Replace
Replace the couch slide servo amplifier
Replace the couch slide servo amplifier — the drive electronics that, together with the gantry KGTSM, powers the longitudinal (IN/OUT) motor M11 (see Related — longitudinal-drive-theory) — by clearing the couch covers that block access, disconnecting the unit's cables, unbolting it from the couch, and then separating the amplifier from its mounting plate on the bench.
Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Basic-to-intermediate — cable disconnection plus M6/M4 fastener work to pull the servo amplifier unit and separate it from its mounting plate; source states no torque value or driver type for either fastener size (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
3 steps
Source
2D201-062E*F p.53-54
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| M6 mounting bolt | n/a in source | 2 |
| M4 screw | n/a in source | count not stated in source |
Full step-by-step for Replace the couch slide servo amplifier
All 3 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table (Longitudinal) guides
Adjust the slide drive belt tension6 stepsAdjust the couch-top supporting rollers (level and sway)4 stepsReplace the horizontal (longitudinal) drive unit5 stepsTroubleshoot: slide stop accuracy, roller ride, rattle/sway, noise, and no in/out motion0 stepsTroubleshoot: couch top will not slide into/out of the gantry opening13 stepsRun the console malfunction-diagnosis program (hwtest)0 steps