Pular para o conteúdo principal
Alarm — high severityRTM REC — recon-process stage

C0020002Reconstruction fails with a recon-process error (RTM REC)

Canon Medical (Toshiba) CT error code, reported by the RTM REC (recon-process stage).

What C0020002 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0020002 indicates: Reconstruction fails with a recon-process error (RTM REC). The code originates in the RTM REC — recon-process stage section of the RTM REC (recon-process stage). Severity level A1 (Alarm — high severity).

Reported action: Try the reconstruction again from the stored raw data and check whether the same series fails a second time, then review the images before releasing the patient. Record the protocol and reconstruction mode and escalate to your service representative — the table carries no further detail for this code..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 101:Device RTM REC group of the console error table, where the OEM text reads "Error in recon-process.", graded A1 (high-severity alarm). RTM is the real time manager on the CT console; REC is the reconstruction stage that turns the stored raw data into images.

Full troubleshooting for C0020002

The complete diagnosis, cause tree, and step-by-step remedy for the RTM REC — recon-process stage subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM REC — recon-process stage codes

Related service procedures

Discuss C0020002 with other field engineers in the FSELIB forum

Source: errmsg.eng — 101:Device RTM REC table. Reference information re-authored by FSELIB — not a copy of the OEM manual.