Information / dialogInjectorSync device

BD210000InjectorSync flags injector abnormal condition

Canon Medical (Toshiba) CT error code, reported by the InjectorSync device.

What BD210000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code BD210000 indicates: InjectorSync flags injector abnormal condition. The code originates in the InjectorSync device section of the InjectorSync device. Severity level A1 (Information / dialog).

Reported action: Confirm the injector is connected, powered and ready, inspect the InjectorSync cable/connector for a loose or damaged link, and clear the fault before the next protocol; if it recurs, cross-check against the injector's own service documentation alongside the CT service manual..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, on the InjectorSync device — the interface that synchronizes the CT scanner with the injector. The same message text is also reported under codes 9D210000, 9D410000, BD110000, BD310000 and BDF10000 in the service error-message table.

Full troubleshooting for BD210000

The complete diagnosis, cause tree, and step-by-step remedy for the InjectorSync device subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related InjectorSync device codes

Related service procedures

Discuss BD210000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.