9D510000Error returned from the injector

Canon Medical (Toshiba) CT error code, reported by the InjectorSync device.

What 9D510000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 9D510000 indicates: Error returned from the injector. The code originates in the InjectorSync device section of the InjectorSync device. Severity level A1 (Information / dialog).

Reported action: Identify the specific fault reported by the injector itself first — the CT console is relaying an error that came back from the injector over the InjectorSync interface — then check that interface cable and connector before restarting the protocol; resolve the reported fault per the injector's own service documentation together with the CT service manual..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems, on the InjectorSync device, when the injector itself returns an error status back to the CT console over that interface.

Full troubleshooting for 9D510000

The complete diagnosis, cause tree, and step-by-step remedy for the InjectorSync device subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related InjectorSync device codes

Related service procedures

Discuss 9D510000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.