Pular para o conteúdo principal
Error conditionX-ray / tube

2127HFG communication error : SCB failure.

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 2127 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 2127 indicates: HFG communication error : SCB failure.. The code originates in the X-ray / tube section of the console error system (EMS). Severity level error (Error condition).

Reported action: X-ray Exposure Cotrol. An error was detected in the x-ray exposure control system. Solution - This may be an intermittent problem and you may be able to continue scanning. If the problem persists call service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.

Full troubleshooting for 2127

The complete diagnosis, cause tree, and step-by-step remedy for the X-ray / tube subsystem are documented in the FSELIB Canon Medical (Toshiba) service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related X-ray / tube codes

Related service procedures

Discuss 2127 with other field engineers in the FSELIB forum

Source: errmsg.eng — Toshiba/Canon CT console (EMS) error message table. Reference information re-authored by FSELIB — not a copy of the OEM manual.