Error conditionX-ray / tube

210072, 102kW:HV Over Current+Arc, 50kW:Arc(+)or Arc(-)

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 2100 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 2100 indicates: 72, 102kW:HV Over Current+Arc, 50kW:Arc(+)or Arc(-). The code originates in the X-ray / tube section of the console error system (EMS). Severity level error (Error condition).

Reported action: X-ray Exposure Cotrol. An error was detected in the x-ray exposure control system, therefore the scan was aborted. Solution - This may be an intermittent problem and you may be able to continue scanning. If the problem persists call service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.