2007 — Rotor startup error : X-ray exposure signal is received when rotor is not ready. It happened when TETD X-ray tube is used.
Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).
What 2007 means
On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 2007 indicates: Rotor startup error : X-ray exposure signal is received when rotor is not ready. It happened when TETD X-ray tube is used.. The code originates in the X-ray / tube section of the console error system (EMS). Severity level error (Error condition).
Reported action: X-ray Tube Not Ready. The x-ray tube is not ready for exposure. Solution - Wait for two minutes and try again. If the problem persists call service..
Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.