Error conditionX-ray / tube

210172, 102kW:Over mA(Fixed)+Abnormal HFXRAY, 50kW:Arc(+) + Arc(-)

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 2101 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 2101 indicates: 72, 102kW:Over mA(Fixed)+Abnormal HFXRAY, 50kW:Arc(+) + Arc(-). The code originates in the X-ray / tube section of the console error system (EMS). Severity level error (Error condition).

Reported action: Tube Arc Detected. An x-ray tube arc was detected, therefore the scan was aborted. Solution - This may be an intermittent problem and you may be able to continue scanning. If the problem persists call service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.