Error conditionX-ray / tube

211272, 102kW:Incorrect exposure detection, 50kW:Abnornal X-ray : SCB exposed X-ray without command from the XC board.

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 2112 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 2112 indicates: 72, 102kW:Incorrect exposure detection, 50kW:Abnornal X-ray : SCB exposed X-ray without command from the XC board.. The code originates in the X-ray / tube section of the console error system (EMS). Severity level error (Error condition).

Reported action: X-ray Exposure Cotrol. An error was detected in the x-ray exposure control system. Solution - This may be an intermittent problem and you may be able to continue scanning. If the problem persists call service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.

Full troubleshooting for 2112