AlarmOutput Sorter Module — Entry/Exit Sensors

HC11240Sorter exit sensor S27 stuck on (continuous activation) at startup

Agfa CT error code, reported by the Output Sorter Module.

What HC11240 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11240 indicates: Sorter exit sensor S27 stuck on (continuous activation) at startup. The code originates in the Output Sorter Module — Entry/Exit Sensors section of the Output Sorter Module. Severity level alarm (Alarm).

Reported action: Check the sorter section and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..

Raised on the Drystar 5500/5503 output sorter module during printer startup/initialization or the empty-check procedure, when sorter exit sensor S27 stays continuously activated because a film remained at its location. Related to HC11239 (entry sensor S25) and HC11241, the dynamic-trigger variant of this same exit sensor.

Full troubleshooting for HC11240

The complete diagnosis, cause tree, and step-by-step remedy for the Output Sorter Module — Entry/Exit Sensors subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual