HC10823 — Sorter exit sensor S27 not connected or malfunctioning
Agfa CT error code, reported by the Output Sorter Module.
What HC10823 means
On DryStar 5500 and DryStar 5503 CT systems, status code HC10823 indicates: Sorter exit sensor S27 not connected or malfunctioning. The code originates in the Output Sorter Module — Position Sensors section of the Output Sorter Module. Severity level alarm (Alarm).
Reported action: Check the connections to S27, then connector P30 on the MEC board, then the wiring between S27 and MEC/P30 with a multimeter, fixing whatever is not OK. If the wiring is good, power off the printer and replace S27; if the fault stays, the MEC board itself has to be replaced..
Raised on the DryStar 5500/5503 output sorter module when the sorter exit sensor S27 is not connected or is malfunctioning. The printer shows "Call service". Filed with the other output-sorter fault codes (HC10823, HC10831-HC10835, HC10907-HC10909, HC11223-HC11224, HC11238-HC11241).