HC11239 — Sorter entry sensor S25 triggered by residual jammed film at startup
Agfa CT error code, reported by the Output Sorter Module.
What HC11239 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11239 indicates: Sorter entry sensor S25 triggered by residual jammed film at startup. The code originates in the Output Sorter Module — Entry/Exit Sensors section of the Output Sorter Module. Severity level alarm (Alarm).
Reported action: Check the sorter section and clear the film jam; if it keeps recurring, look for mechanical blockage of the sheet transport and for broken or misaligned sensors..
Raised on the Drystar 5500/5503 output sorter module during printer startup/initialization or the empty-check procedure, when sorter entry sensor S25 is triggered by a film jammed and still sitting in the sorter section. Its counterparts are HC11240 (exit sensor S27 stuck on) and HC11241 (exit sensor S27 triggered).
Full troubleshooting for HC11239
The complete diagnosis, cause tree, and step-by-step remedy for the Output Sorter Module — Entry/Exit Sensors subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Output Sorter Module — Entry/Exit Sensors codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.