HC10831 — Sorter entry sensor S25 not connected or malfunctioning
Agfa CT error code, reported by the Output Sorter Module.
What HC10831 means
On DryStar 5500 and DryStar 5503 CT systems, status code HC10831 indicates: Sorter entry sensor S25 not connected or malfunctioning. The code originates in the Output Sorter Module — Position Sensors section of the Output Sorter Module. Severity level alarm (Alarm).
Reported action: Check the connections to S25, then connector P30 on the MEC board, then the wiring between S25 and MEC/P30 with a multimeter, fixing whatever is not OK. If the wiring is good, power off the printer and replace S25; if the fault stays, the MEC board itself has to be replaced..
Raised on the DryStar 5500/5503 output sorter module when the sorter entry sensor S25 is not connected or is malfunctioning. The printer shows "Call service". Filed with the other output-sorter fault codes (HC10823, HC10831-HC10835, HC10907-HC10909, HC11223-HC11224, HC11238-HC11241).
Full troubleshooting for HC10831
The complete diagnosis, cause tree, and step-by-step remedy for the Output Sorter Module — Position Sensors subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Output Sorter Module — Position Sensors codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.