Replace the inverter in the SRU generator

Replace an individual inverter in the SRU generator — and its matching filter-board fuse — by removing the top cover chassis, disconnecting the four input/signal cables, unbolting the filter screws to lift the assembly clear, then freeing the radiator so the inverter itself can be swapped.

Applies to
Aquilion 16 (TSX-101A)
Personnel
Intermediate — reach the inverter by removing the SRU generator's top cover chassis (keeping the near-side drive power supply mounted on it), disconnecting the four input/signal cables, unbolting the screws at J1 to J10(11) and J14 to J18 on both filters to lift the assembly clear together with the drive power supply, then freeing the screws at both ends of the inverter's radiator — ideally with a magnetic screwdriver, since a dropped screw is otherwise hard to recover from around the radiator — before swapping the inverter and its matching filter-board fuse together. No torque value, electrical test, or calibration step is stated within this section itself. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
5 steps
Source
2D201-064E*H p.117

Tools & parts

ItemPart numberQty
Magnetic screwdrivern/a in source1 — optional; source states the radiator may instead be lifted out together with its screws if one is not available
Replacement invertern/a in sourcen/a in source
Corresponding filter-board fusen/a in sourcen/a in source

Full step-by-step for Replace the inverter in the SRU generator

All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related X-ray & HV (SRU generator) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-064E*H p.117. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.