Replace the 4-MHU X-ray tube starter
Remove and replace the 4-MHU X-ray tube starter chassis on the Activion16 gantry, then set the microcomputer circuit's tube-type DIP switch for the installed tube before remounting. This is a compact FRU swap: unbolt the starter from the gantry, disconnect its cabling, reconfigure it for the tube in use, and reinstall with the mounting screws torqued and witness-marked.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| 4-MHU X-ray tube starter (chassis assembly) | ST-7013 (per source figure caption — see naming note below) | 1 |
Safety
Hazard: Changing DIP switch S1 on the microcomputer circuit while the starter is powered
Consequence: Electric shock
Avoidance: Confirm the starter is de-energized (system power OFF at the distribution board) before touching DIP switch S1; never change S1 with power applied
Full step-by-step for Replace the 4-MHU X-ray tube starter
All 5 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV guides
Source: 2D201-112EN*D §1.4.4 (pp.71–72). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.