Remove and replace the 4-MHU X-ray tube starter chassis on the Activion16 gantry, then set the microcomputer circuit's tube-type DIP switch for the installed tube before remounting. This is a compact FRU swap: unbolt the starter from the gantry, disconnect its cabling, reconfigure it for the tube in use, and reinstall with the mounting screws torqued and witness-marked.

Applies to
Activion16 (TSX-031A) with the 4-MHU X-ray tube starter (ST-7013) and 42-kW HFG configuration, per the source figure.
Personnel
1 technician · Field service engineer (CT gantry / X-ray & HV)
Steps
5 steps
Source
2D201-112EN*D §1.4.4 (pp.71–72)

Tools & parts

ItemPart numberQty
4-MHU X-ray tube starter (chassis assembly)ST-7013 (per source figure caption — see naming note below)1

Safety

DANGER

Hazard: Changing DIP switch S1 on the microcomputer circuit while the starter is powered

Consequence: Electric shock

Avoidance: Confirm the starter is de-energized (system power OFF at the distribution board) before touching DIP switch S1; never change S1 with power applied

Full step-by-step for Replace the 4-MHU X-ray tube starter

All 5 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.