Pular para o conteúdo principal
ProcedureX-ray & HVReplacementReplace

Replace the 42-kW HFG (CXXG-010A)

Replace one of the two field-replaceable chassis of the 42-kW X-ray high-frequency generator (HFG) mounted on the rotation base: the AC/CONTROL chassis (Part 1) or the HV/INVERTER CATHODE chassis (Part 2). Both are swapped the same way — index the rotation section under the X-ray tube replacement jig, rig the chassis on a chain block, unbolt it, and mount the replacement — but each is its own inde

Applies to
Activion16 (TSX-031A) systems equipped with the 42-kW HFG assembly (CXXG-010A). No further model/revision restriction is stated in source.
Personnel
2 technicians · Field service engineer (CT gantry / HV, chain-block rigging)
Steps
43 steps
Source
2D201-112EN*D §1.4.7, §1.4.7.1, §1.4.7.2 (pp.78–82)

Tools & parts

ItemPart numberQty
42-kW HFG assemblyCXXG-010A1
AC/CONTROL chassis (new)n/a in source — see IPD1
HV/INVERTER CATHODE chassis (new)n/a in source — see IPD1

Safety

WARNING

Hazard: Gantry front cover left unsupported, or lifted inadvertently, after opening

Consequence: Supporting poles fall and the cover closes suddenly under its own weight — injury

Avoidance: Support the open front cover with two supporting poles at the rear; use the poles correctly

WARNING

Hazard: Removing a rotation-section part (AC/CONTROL or HV/INVERTER CATHODE chassis) without the rotation mechanical stopper engaged

Consequence: Weight imbalance causes the rotation section to rotate suddenly, catching and seriously injuring a worker

Avoidance: Engage the rotation mechanical stopper of the rotation drive section before removing the part; hold the rotation section firmly while pulling out the stopper

CAUTION

Hazard: Wire or chain block used to rig the chassis is damaged or otherwise unsound

Consequence: Damage to the system or units if the suspended chassis drops or swings uncontrolled

Avoidance: Confirm there are no problems with the wire and chain block before loading them; hold the chassis firmly while removing its securing bolts

Full step-by-step for Replace the 42-kW HFG (CXXG-010A)

All 43 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related X-ray & HV guides

View all Activion16 (TSX-031A) guides

Source: 2D201-112EN*D §1.4.7, §1.4.7.1, §1.4.7.2 (pp.78–82). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.