Skip to main content
ProcedureX-ray & HVReplacementReplace

Replace the X-ray tube

Remove and replace the X-ray tube (and its oil cooler) on the Activion16 rotation base, using the tube replacement jig and a chain block. The tube and oil cooler are lowered as suspended loads while the rotation section is manually indexed under the jig.

Applies to
Activion16 (TSX-031A) with 4-MHU X-ray tube and ratchet chain-block tube jig CLX-21141-1.
Personnel
2 technicians · Field service engineer (CT gantry / HV)
Steps
25 steps
Source
2D201-112EN*D §1.4.3 (pp.63–70)

Tools & parts

ItemPart numberQty
Tube replacement jigCLX-21141-11 set
X-ray tube (4-MHU)<PN — see IPD>1

Safety

DANGER

Hazard: Residual high-voltage charge in the HV cable / bushing

Consequence: Electric shock

Avoidance: Power OFF ≥5 min; ground 2 of the 3 bushing pins in all 3 combinations twice or more; never touch the bushing tip; cap the tip

WARNING

Hazard: Unbalanced rotation section rotates suddenly

Consequence: Worker caught in rotating section — serious injury

Avoidance: Engage the rotation mechanical stopper; hold the rotation section firmly when pulling the stopper

WARNING

Hazard: Unsupported gantry front cover closes under its own weight

Consequence: Injury

Avoidance: Support the open cover with two supporting poles at the rear; use the poles correctly

WARNING

Hazard: Energizing the rotation unit during service

Consequence: Unexpected rotation — serious injury

Avoidance: Never turn ON breaker NFB2 of the rotation unit

CAUTION

Hazard: Tube moves longitudinally while mounting bolts are out

Consequence: Pinch / dropped-load

Avoidance: Steady the tube; keep hands clear of travel path

CAUTION

Hazard: Suspended tube / oil cooler on chain block

Consequence: Dropped load — equipment damage / injury

Avoidance: Verify wire and chain block condition; lock the jig brake before lowering

Full step-by-step for Replace the X-ray tube

All 25 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

7-day free trial · no credit card required

Related X-ray & HV guides

View all Activion16 (TSX-031A) guides

Related error codes

Source: 2D201-112EN*D §1.4.3 (pp.63–70). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.

“That they may see, and know, and consider, and understand together, that the hand of the LORD hath done this.”Isaiah 41:20