Adjust the X-ray tube alignment (front/rear and left/right)

Realign the X-ray tube on the rotation base after tube replacement or whenever beam position drifts out of tolerance. The task has two independent axes — front/rear and left/right — each measured through the alignment utility in the service manual's software volume and corrected mechanically with a dial gauge and the tube mounting plate's adjusting bolts. Do front/rear first, then left/right.

Applies to
Activion16 (TSX-031A). No model/revision restriction is stated in source for this section. Typically performed as the direct follow-on to X-ray tube replacement (§1.4.3) or whenever tube position drifts out of tolerance.
Personnel
1 technician · "Field service engineer (CT gantry mechanical service), with access to the tube-alignment utility in the service manual's software volume"
Steps
13 steps
Source
2D201-112EN*D §1.5.3 (pp.90–93) — Tube alignment (1.5.3.1 Outline, 1.5.3.2 Front/rear, 1.5.3.3 Left/right)

Tools & parts

ItemPart numberQty
Torque wrench—1
Yellow paint pen—1
Dial gauge—1
Pin phantom—1
Phantom holder—1

Full step-by-step for Adjust the X-ray tube alignment (front/rear and left/right)

All 13 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.