Pular para o conteúdo principal
Alarm — high severityRDD (Raw Data Disk Array)

C003000DBad disk detected while configuring the RDD RAID

Canon Medical (Toshiba) CT error code, reported by the RDD (Raw Data Disk Array) — RAID status, reported via RTM.

What C003000D means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C003000D indicates: Bad disk detected while configuring the RDD RAID. The code originates in the RDD (Raw Data Disk Array) section of the RDD (Raw Data Disk Array) — RAID status, reported via RTM. Severity level A1 (Alarm — high severity).

Reported action: A bad drive was found during RAID configuration. Identify it from the per-slot LEDs on the RDD front panel, replace it using the disk-replacement procedure, then re-run RDD Setup and reconfigure the RDD with parity..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, on the RDD (Raw Data Disk Array) — the fibre-channel raw-data disk array, run as a RAID set with parity, that stores the raw scan data preprocessed by the MHR boards before reconstruction — reported under the RTM (real time manager) RDD device group of the system error message table.

Full troubleshooting for C003000D

The complete diagnosis, cause tree, and step-by-step remedy for the RDD (Raw Data Disk Array) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RDD (Raw Data Disk Array) codes

Related service procedures

Discuss C003000D with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, set 102 — Device RTM RDD): "Error in RDD. (Bad disk detected when configuring RAID.)". Reference information re-authored by FSELIB — not a copy of the OEM manual.