C003000A — RDD RAID cannot run with disks of different types
Canon Medical (Toshiba) CT error code, reported by the RDD (Raw Data Disk Array) — RAID status, reported via RTM.
What C003000A means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C003000A indicates: RDD RAID cannot run with disks of different types. The code originates in the RDD (Raw Data Disk Array) section of the RDD (Raw Data Disk Array) — RAID status, reported via RTM. Severity level A1 (Alarm — high severity).
Reported action: The RAID will not run with mixed disk types. Verify that every drive fitted in the RDD is the correct matching type for that array, replace any odd drive using the disk-replacement procedure, and re-run RDD Setup afterwards..
Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, on the RDD (Raw Data Disk Array) — the fibre-channel raw-data disk array, run as a RAID set with parity, that stores the raw scan data preprocessed by the MHR boards before reconstruction — reported under the RTM (real time manager) RDD device group of the system error message table.
Full troubleshooting for C003000A
The complete diagnosis, cause tree, and step-by-step remedy for the RDD (Raw Data Disk Array) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated RDD (Raw Data Disk Array) codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng, set 102 — Device RTM RDD): "Error in RDD. (RAID cannot run with disks of different types)". Reference information re-authored by FSELIB — not a copy of the OEM manual.