B7110004 — Channel-switching error in the MUDAT data transfer link
Canon Medical (Toshiba) CT error code, reported by the MUDAT (Data Transfer).
What B7110004 means
On Aquilion RX / V4-62R004 and Aquilion ONE / V4.2 CT systems, status code B7110004 indicates: Channel-switching error in the MUDAT data transfer link. The code originates in the MUDAT (gantry-to-console data transfer) section of the MUDAT (Data Transfer). Severity level A1 (Alarm — high severity).
Reported action: Check the MUDAT channel path end to end — the interface boards on the rotation and stationary sides and their connectors — then run the MUDAT test procedure in the service manual to identify which channel fails. Turn the gantry off for a few seconds and retry before replacing any board..
Toshiba/Canon Aquilion RX (V4-62R004) and Aquilion ONE (V4.2) CT systems, on the MUDAT link, where acquisition data is carried between the gantry rotation section and the stationary section over several transfer channels. This code is reported when switching between those MUDAT channels fails.
Full troubleshooting for B7110004
The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (gantry-to-console data transfer) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated MUDAT (gantry-to-console data transfer) codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng, Aquilion RX V4-62R004 / Aquilion ONE V4.2). Reference information re-authored by FSELIB — not a copy of the OEM manual.