Alarm — high severityMUDAT (gantry-to-console data transfer)

B7110002Data transfer system (1) error during scanning (MUDAT)

Canon Medical (Toshiba) CT error code, reported by the MUDAT (Data Transfer).

What B7110002 means

On Aquilion RX / V4-62R004 and Aquilion ONE / V4.2 CT systems, status code B7110002 indicates: Data transfer system (1) error during scanning (MUDAT). The code originates in the MUDAT (gantry-to-console data transfer) section of the MUDAT (Data Transfer). Severity level A1 (Alarm — high severity).

Reported action: First check the images from the affected scan for streak artifacts before they are used clinically — the message itself asks the operator to confirm the images are free of abnormalities. Then inspect data transfer system (1) on the MUDAT link (interface boards and connector seating on the rotation and stationary sides) and repeat the scan; if the error returns, isolate the link with the MUDAT test procedure in the service manual..

Toshiba/Canon Aquilion RX (V4-62R004) and Aquilion ONE (V4.2) CT systems, on the MUDAT link that moves acquisition data from the gantry rotation section to the stationary/console section. This code is raised when data transfer system (1) fails while a scan is actually running, so the images from that scan can be affected.

Full troubleshooting for B7110002

The complete diagnosis, cause tree, and step-by-step remedy for the MUDAT (gantry-to-console data transfer) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related MUDAT (gantry-to-console data transfer) codes

Related service procedures

Discuss B7110002 with other field engineers in the FSELIB forum