ProcedureX-ray TubeReplacement/FRUReplace

Connect the replacement X-ray tube and run the operation check

After the X-ray tube itself has been mechanically replaced (Volume 1 procedure), complete the electrical connection and operation check described here before returning the system to service: rewire the tube leads to the starter terminal block, confirm the XCS PWB tube-type setting, verify basic operation, then work through the If-adjustment / seasoning / alignment follow-on tasks.

Applies to
Activion16 (TSX-031A). Electrical-side connection and operation-check sequence run after the mechanical X-ray tube swap (Volume 1 procedure); no model/revision restriction is stated in source for this section. The tube-lead-to-TBS01-pin correspondence is preserved exactly as the source's two-column layout extracts it; the extraction does not clearly pair each lead color to a specific pin number (see FSELIB note in Step group A).
Personnel
Field service engineer (CT X-ray tube / X-ray generator electrical connection & checkout)
Steps
3 steps
Source
2D201-113EN*J §1.10 'Replacing the X-ray Tube' preamble (p.66) + §1.10.1 'Connection and operation check' (p.67)

Tools & parts

ItemPart numberQty
XCS PWB (in the AC/CONTROL unit)n/a in source1
X-ray tube unit (stationary unit) leads — White / Black / Orangen/a in source1
Starter terminal block TBS01 (pins 1 Red / 2 White / 3 Black)n/a in source1

Safety

NOTICE

Hazard: Connecting the starter cables at X-ray tube replacement without checking the cable marker against the circuit diagrams

Consequence: Damage to the anode rotation mechanism

Avoidance: Refer to the circuit diagrams and check the marker placed on the cable to perform correct cable connection

Full step-by-step for Connect the replacement X-ray tube and run the operation check

All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related X-ray Tube guides

View all Activion16 (TSX-031A) guides

Source: 2D201-113EN*J §1.10 'Replacing the X-ray Tube' preamble (p.66) + §1.10.1 'Connection and operation check' (p.67). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.