Clean the heat exchanger fan (annual PM)

If the heat exchanger's radiator is clogged with dust, the fan's air volume drops and cooling efficacy is reduced. Clean the heat exchanger fan and radiator on the gantry-mounted oil cooler at least once a year to prevent this.

Applies to
Activion16 (TSX-031A) X-ray tube unit / heat exchanger (CXB-400C tube system). Calendar-based preventive-maintenance functional check: clean the heat exchanger's radiator and fan on the gantry-mounted oil cooler. Cleaning interval per source is at least once a year. Source does not state any model- or revision-specific restriction for this section.
Steps
7 steps
Source
2D201-113EN*J §1.11.5 'Heat exchanger fan cleaning' (pp.109-110)

Tools & parts

ItemPart numberQty
Oil cooler unit (heat exchanger: radiator + fan)n/a in source1
Oil cooler cover screwsn/a in source10
Vacuum cleanern/a in source1

Full step-by-step for Clean the heat exchanger fan (annual PM)