Replace the REF detector
The REF (reference) detector sits behind the leakage beam shielding plate. On systems with the small-focus X-ray tube (CXB-400C), reaching it means first clearing the optical system shielding cover, the optical system (slit/wedge), and the X-ray tube out of the way. The REF detector itself is a high-precision part — handle it with care throughout.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Standard tools | n/a in source | 1 set |
| Multimeter | n/a in source | 1 |
| Insulated cable ties | n/a in source | Several |
| REF detector (replacement unit) | n/a in source | 1 |
| Insulating washers (REF detector mount) | n/a in source | n/a in source |
Safety
Hazard: Touching the optical-system shield-cover drive motor while removing the front cover
Consequence: Burn injury — the motor is hot
Avoidance: Do not touch this motor
Hazard: Mishandling the REF detector during removal or remounting (it is a high-precision part)
Consequence: Damage to the REF detector (equipment damage only — no injury is stated in source for this item)
Avoidance: Handle the REF detector with care throughout removal and remounting
Full step-by-step for Replace the REF detector
All 11 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Detector guides