Perform Print Head (TH) Calibration
Recalibrate the DryStar 5300's thermal print head (TPH) profile whenever vertical stripes become visible in the medical service test image — after the TH cleaning procedure has already been run, since this calibration only compensates macro-TPH banding wider than 10-15 mm, never sharp-edged artifacts.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Calibrated densitometer — MacBeth TR-924, X-Rite 331, X-Rite 341, X-Rite 301, X-Rite 310, or Gretag D200-2 | n/a in source | |
| Ink eraser 'Pelikan' | 10+999991176.0 |
Full step-by-step for Perform Print Head (TH) Calibration
All 5 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Calibration guides
Source: DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6 §2.3 Print Head (TH) Calibration, p.375-377 (chapter-relative footers "Chapter 3.6 / 7" through "Chapter 3.6 / 9"), Edition 1, Revision 7 (2006-01-26). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.