Perform CDM (internal densitometer) calibration
Recalibrate the DryStar 5300's internal densitometer (CDM) against an external calibrated densitometer — from the Service Command Session webpages — whenever a post-film-calibration density reproduction check comes back not OK, or after the Output module (which contains the CDM) has been replaced.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| External calibrated densitometer | n/a in source | 1 |
Full step-by-step for Perform CDM (internal densitometer) calibration
All 7 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.