Perform CDM (internal densitometer) calibration

Recalibrate the DryStar 5300's internal densitometer (CDM) against an external calibrated densitometer — from the Service Command Session webpages — whenever a post-film-calibration density reproduction check comes back not OK, or after the Output module (which contains the CDM) has been replaced.

Applies to
DryStar 5300, Type 5365/100 (source Chapter 3.6 §2.2 'CDM calibration', Repair and Service: Adjustments and calibrations, Edition 1, Revision 7, 2006-01-26) — calibrates the internal densitometer (CDM) against an external calibrated densitometer, via the Service Command Session webpages.
Personnel
1 technician · Service-level technician — navigates the Service Engineer Tools > Enter Service Command session webpages, operates and reads an external calibrated densitometer, and enters 20 measured density readings (steps 00-19) into the browser form. No disassembly is described in this section (derived — source does not assign a formal skill-level label to this section).
Steps
7 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6/5-6, §2.2 'CDM calibration', Edition 1, Revision 7 (2006-01-26), document p.373-374

Tools & parts

ItemPart numberQty
External calibrated densitometern/a in source1

Full step-by-step for Perform CDM (internal densitometer) calibration

All 7 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Calibration guides

View all DryStar 5300 guides

Source: DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6/5-6, §2.2 'CDM calibration', Edition 1, Revision 7 (2006-01-26), document p.373-374. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.