Perform Film Calibration

Run the DryStar 5300's film calibration whenever the printed Dmax has drifted, after a Thermal print head (TPH) swap, or simply on the automatic schedule — then read the result against the target-Dmax table for the densitometer configured on the printer.

Applies to
DryStar 5300, Type 5365/100 (source Chapter 3.6, Repair and Service: Adjustments and calibrations, Edition 1, Revision 7, 2006-01-26) — §2.1 Film Calibration.
Personnel
1 technician · Basic — start the calibration from the Key-operator menu, read the printed calibration film's measured Dmax against the target-Dmax table for the configured densitometer, and recognize the pass / auto-retry / "calibration failed" outcomes. No disassembly or special test equipment; an external densitometer is only needed if you go on to Troubleshooting (derived — source does not assign a formal skill-level label to this section).
Steps
4 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6 §2.1 Film Calibration, p.371-372 (chapter-relative footers "Chapter 3.6 / 3" and "Chapter 3.6 / 4"), Edition 1, Revision 7 (2006-01-26)

Full step-by-step for Perform Film Calibration

All 4 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Calibration guides

View all DryStar 5300 guides

Source: DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6 §2.1 Film Calibration, p.371-372 (chapter-relative footers "Chapter 3.6 / 3" and "Chapter 3.6 / 4"), Edition 1, Revision 7 (2006-01-26). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.