Perform density reproduction check
Run the density reproduction check on every film type the printer prints, so the pass/fail readout on the worksheet — not a guess — decides whether the CDM needs a re-calibration.
Applies to
DryStar 5300 (Type 5365/100). Must be carried out with every film type in use (source, chapter 3.6/11).
Personnel
1 technician · Service-level technician — logs into the printer's browser-based Service Engineer Tools, prints a service test image, and takes 11 manual density readings with a calibrated external densitometer, then transcribes the values into a MedNet-supplied Excel worksheet that computes the pass/fail verdict automatically. No disassembly and no test equipment beyond the densitometer itself. (derived — source does not assign a formal skill-level label to this section.)
Steps
5 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6/10-12, §3.1 'Density reproduction check', Edition 1, Revision 7, 2006-01-26, Type 5365/100
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Calibrated densitometer — MacBeth TR-924, X-Rite 331, X-Rite 341, X-Rite 301, X-Rite 310, or Gretag D200-2 | n/a in source | 1 |
| Test image `LinCheck.zip` | n/a in source | 1 print |
| Excel worksheet "Ds5300 density reproduction check DT2B.xls" (blue-based film) or "...DT2C.xls" (clear-based film) | n/a in source | 1, matching the film type under test |
Full step-by-step for Perform density reproduction check
All 5 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Calibration guides
Perform CDM (internal densitometer) calibration7 stepsCheck for dust stripes and perform TPH cleaning procedure7 stepsPerforming the Film Calibration Procedure3 stepsPerform Film Calibration4 stepsPerform horizontal film alignment procedure7 stepsPerform Print Head (TH) Calibration5 steps
View all DryStar 5300 guidesSource: DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6/10-12, §3.1 'Density reproduction check', Edition 1, Revision 7, 2006-01-26, Type 5365/100. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.