Perform density reproduction check

Run the density reproduction check on every film type the printer prints, so the pass/fail readout on the worksheet — not a guess — decides whether the CDM needs a re-calibration.

Applies to
DryStar 5300 (Type 5365/100). Must be carried out with every film type in use (source, chapter 3.6/11).
Personnel
1 technician · Service-level technician — logs into the printer's browser-based Service Engineer Tools, prints a service test image, and takes 11 manual density readings with a calibrated external densitometer, then transcribes the values into a MedNet-supplied Excel worksheet that computes the pass/fail verdict automatically. No disassembly and no test equipment beyond the densitometer itself. (derived — source does not assign a formal skill-level label to this section.)
Steps
5 steps
Source
DD+DIS346.05E — Drystar 5300 Repair and Service, Adjustments and calibrations, Chapter 3.6/10-12, §3.1 'Density reproduction check', Edition 1, Revision 7, 2006-01-26, Type 5365/100

Tools & parts

ItemPart numberQty
Calibrated densitometer — MacBeth TR-924, X-Rite 331, X-Rite 341, X-Rite 301, X-Rite 310, or Gretag D200-2n/a in source1
Test image `LinCheck.zip`n/a in source1 print
Excel worksheet "Ds5300 density reproduction check DT2B.xls" (blue-based film) or "...DT2C.xls" (clear-based film)n/a in source1, matching the film type under test

Full step-by-step for Perform density reproduction check

All 5 steps are documented in the FSELIB DryStar 5300 service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure