ER26 — Stator Control Board to Data Receiving Board communication failure
Philips CT error code, reported by the Data Path (Slip Ring / Image Data Transfer).
What ER26 means
On Access Series CT (6/16/Dual) CT systems, status code ER26 indicates: Stator Control Board to Data Receiving Board communication failure. The code originates in the Data Path (Image Data Transfer) section of the Data Path (Slip Ring / Image Data Transfer). Severity level alarm (Alarm — stator control board error condition).
Reported action: Cross-reference System Error Code 0x5717 in the system log. Use the Data Path Troubleshooting flow (Section 2.6.2) and check the link between the Stator Control Board and the DRC/Data Receiving board before replacement..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.
Full troubleshooting for ER26
The complete diagnosis, cause tree, and step-by-step remedy for the Data Path (Image Data Transfer) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Data Path (Image Data Transfer) codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.