ER1E — Hotlink chip losing lock
Philips CT error code, reported by the Data Path (Slip Ring / Image Data Transfer).
What ER1E means
On Access Series CT (6/16/Dual) CT systems, status code ER1E indicates: Hotlink chip losing lock. The code originates in the Data Path (Image Data Transfer) section of the Data Path (Slip Ring / Image Data Transfer). Severity level alarm (Alarm — stator control board error condition).
Reported action: Cross-reference System Error Code 0x5709 in the system log. Use the Data Path Overview/Troubleshooting flow (Section 2.6) and check the optical/serial link and connectors around the Hotlink transceiver before board replacement..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.