Pular para o conteúdo principal
Warning — recoverable conditionHost computer (HP Workstation x2100)

0012HPC configuration lost (CMOS and SE2P backup both lost)

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0012H means

On Infinion 1.5T CT systems, status code 0012H indicates: PC configuration lost (CMOS and SE2P backup both lost). The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level warning (Warning — recoverable condition).

Reported action: Enter BIOS Setup, restore default settings, and reconfigure the workstation (boot device order, integrated ports, etc.). Recurrence after every power-down indicates a system-board CMOS/SE2P backup fault requiring board service..

Documented as EMU code 0012h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "PC configuration lost (both SE2P and CMOS lost)"; short message: "Incorrect PC configuration". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0012H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0012H with other field engineers in the FSELIB forum

Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.