Pular para o conteúdo principal
Warning — recoverable conditionHost computer (HP Workstation x2100)

0041Invalid or uninitialized internal product-type flag

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0041 means

On Infinion 1.5T CT systems, status code 0041 indicates: Invalid or uninitialized internal product-type flag. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level warning (Warning — recoverable condition).

Reported action: The internal product identification flag in system-board storage is uninitialized or corrupted. Contact HP/Philips service support to have the product type reprogrammed via BIOS Setup; this does not by itself block normal operation..

Documented as EMU code 0041 (printed without the trailing 'h' seen on neighboring codes in this row of the source table) in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Product flag not initialized or bad"; short message: "Invalid internal product type". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0041

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0041 with other field engineers in the FSELIB forum

Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.