0011H — System date and time lost from CMOS/SE2P backup
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0011H means
On Infinion 1.5T CT systems, status code 0011H indicates: System date and time lost from CMOS/SE2P backup. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level warning (Warning — recoverable condition).
Reported action: Enter BIOS Setup and re-enter the correct system date and time. If the fault recurs after every power cycle, the onboard battery or SE2P backup circuit on the system board likely needs replacement..
Documented as EMU code 0011h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Date and time (CMOS backed up from SE2P)"; short message: "Date and time lost". Not confirmed by this document for other Philips systems.
Full troubleshooting for 0011H
The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Host computer (HP Workstation x2100) codes
Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.