E171F — System BIOS reported a PCIe fatal error on a component in a specific PCIe slot/bus/device/function
Philips CT error code, reported by the MND Extended Storage Server (Dell PowerEdge 2900 hardware).
What E171F means
On Mobile Network Device (MND) Extended Storage Option CT systems, status code E171F indicates: System BIOS reported a PCIe fatal error on a component in a specific PCIe slot/bus/device/function. The code originates in the System Bus / PCI-PCIe I/O section of the MND Extended Storage Server (Dell PowerEdge 2900 hardware). Severity level alarm-high (Alarm — critical, fatal PCIe error).
Reported action: Remove and reseat the PCI expansion cards (see Troubleshooting Expansion Cards, p.81). If the problem persists, the system board is faulty — see Getting Help (p.10)..
Listed as an LCD front-panel status/error code ('PCIE Fatal Err B## D## F##' / 'PCIE Fatal Err Slot #', Table 2 LCD Status Messages, p.23). Identifies the PCIe bus/device/function or slot.
Full troubleshooting for E171F
The complete diagnosis, cause tree, and step-by-step remedy for the System Bus / PCI-PCIe I/O subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated System Bus / PCI-PCIe I/O codes
Source: Philips Mobile Network Device (MND) Extended Storage Option Troubleshooting Guide, Rev A — 4598 004 17081 (LCD Status Messages table). Reference information re-authored by FSELIB — not a copy of the OEM manual.