200011E2 — Temperature limits exceeded
Philips CT error code, reported by the Thermal.
What 200011E2 means
On CT systems, status code 200011E2 indicates: Temperature limits exceeded. The code originates in the Thermal section of the Thermal. Severity level alarm-high (High alarm — over temperature).
Reported action: Treat as urgent: check cabinet and room cooling — fans running, air filters clean, vents and intakes not blocked — and review the temperature readings for the subsystem in the prefix. Let the system cool and correct the airflow or cooling fault before resuming scanning..
From the Philips ultrasound service reference "Error Logger Overview" (New Error Logger), archived under Philips ultrasound End-of-Life service documentation. The source lists this Errno as "Temperature limits exceeded". A full log entry combines a 4-digit Subsystem code with the Errno (format ssss-eeeeeeee), and the prefix indicates which area went over temperature.
Full troubleshooting for 200011E2
The complete diagnosis, cause tree, and step-by-step remedy for the Thermal subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Thermal codes
Source: Philips Ultrasound (End of Life) — 10_Error Logger Overview, Reference documentation. Reference information re-authored by FSELIB — not a copy of the OEM manual.