0801H — System cache self-test failure
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0801H means
On Infinion 1.5T CT systems, status code 0801H indicates: System cache self-test failure. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).
Reported action: Check that the processor is correctly seated in its socket; if the cache self-test continues to fail, replace the processor per the Processors (Socket) parts list..
Documented as EMU code 0801h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Cache self-test failure"; short message: "System cache error". Not confirmed by this document for other Philips systems.
Full troubleshooting for 0801H
The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Host computer (HP Workstation x2100) codes
Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.