0700H — System memory decreased since previous boot
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0700H means
On Infinion 1.5T CT systems, status code 0700H indicates: System memory decreased since previous boot. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).
Reported action: Check that all RDRAM modules are correctly seated and that RIMM continuity modules are installed in unused sockets; reseat or replace any module reporting a lower size than before..
Documented as EMU code 0700h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Found less DRAM memory than at previous boot"; short message: "System memory error". Not confirmed by this document for other Philips systems.