Pular para o conteúdo principal
Alarm — hardware/component faultHost computer (HP Workstation x2100)

0700HSystem memory decreased since previous boot

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0700H means

On Infinion 1.5T CT systems, status code 0700H indicates: System memory decreased since previous boot. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).

Reported action: Check that all RDRAM modules are correctly seated and that RIMM continuity modules are installed in unused sockets; reseat or replace any module reporting a lower size than before..

Documented as EMU code 0700h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Found less DRAM memory than at previous boot"; short message: "System memory error". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0700H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0700H with other field engineers in the FSELIB forum

Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.