Pular para o conteúdo principal
Alarm — hardware/component faultHost computer (HP Workstation x2100)

0600HVideo memory mismatch (less than configured in CMOS)

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0600H means

On Infinion 1.5T CT systems, status code 0600H indicates: Video memory mismatch (less than configured in CMOS). The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).

Reported action: Check that the AGP video card is correctly seated in its slot and that the BIOS Setup AGP/video configuration matches the installed card; reseat or replace the video card if the mismatch persists..

0600h

Full troubleshooting for 0600H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0600H with other field engineers in the FSELIB forum