Alarm — hardware/component faultHost computer (HP Workstation x2100)

0531HIDE drive found on slave connector only, secondary channel

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0531H means

On Infinion 1.5T CT systems, status code 0531H indicates: IDE drive found on slave connector only, secondary channel. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).

Reported action: Check the secondary IDE cable and jumper/cable-select settings; verify master/slave configuration and connector seating on the secondary channel..

Documented as EMU code 0531h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Found a drive on slave connector only (secondary)"; short message: "IDE device error". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0531H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0531H with other field engineers in the FSELIB forum