0530H — IDE drive found on slave connector only, primary channel
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0530H means
On Infinion 1.5T CT systems, status code 0530H indicates: IDE drive found on slave connector only, primary channel. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).
Reported action: Check the primary IDE cable and jumper/cable-select settings; a drive present only in the slave position with no master drive can cause detection issues. Verify master/slave configuration and connector seating..
Documented as EMU code 0530h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Found a drive on slave connector only (primary)"; short message: "IDE device error". Not confirmed by this document for other Philips systems.
Full troubleshooting for 0530H
The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Host computer (HP Workstation x2100) codes
Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.