Pular para o conteúdo principal
Alarm — hardware/component faultHost computer (HP Workstation x2100)

0522HHard disk drive 2 not detected but configured in CMOS

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0522H means

On Infinion 1.5T CT systems, status code 0522H indicates: Hard disk drive 2 not detected but configured in CMOS. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).

Reported action: Check that drive 2 is correctly installed and cabled, or update the BIOS Setup HDD configuration (use Auto detection) if the drive has been removed..

Documented as EMU code 0522h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "HDD # 2 not detected (but configured in CMOS)"; short message: "IDE device # 2 error". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0522H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0522H with other field engineers in the FSELIB forum