0520H — Hard disk drive 0 not detected but configured in CMOS
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0520H means
On Infinion 1.5T CT systems, status code 0520H indicates: Hard disk drive 0 not detected but configured in CMOS. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).
Reported action: Check that drive 0 is correctly installed and cabled to the primary IDE channel, or update the BIOS Setup HDD configuration (use Auto detection) if the drive has been removed..
Documented as EMU code 0520h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "HDD # 0 not detected (but configured in CMOS)"; short message: "IDE device # 0 error". Not confirmed by this document for other Philips systems.
Full troubleshooting for 0520H
The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Host computer (HP Workstation x2100) codes
Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.