0306H — General floppy disk controller failure
Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).
What 0306H means
On Infinion 1.5T CT systems, status code 0306H indicates: General floppy disk controller failure. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level alarm (Alarm — hardware/component fault).
Reported action: Check the FDD IDE cable (P/N 5184-1516) and its connection to the system board; if the controller continues to fail, the system board may require replacement..
Documented as EMU code 0306h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "General failure on floppy controller"; short message: "Flexible disk drive error". Not confirmed by this document for other Philips systems.
Full troubleshooting for 0306H
The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Host computer (HP Workstation x2100) codes
Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.