Pular para o conteúdo principal
Warning — recoverable conditionHost computer (HP Workstation x2100)

0101HKeyboard self-test failure at POST

Philips CT error code, reported by the Host Computer (HP Workstation x2100 EMU/BIOS).

What 0101H means

On Infinion 1.5T CT systems, status code 0101H indicates: Keyboard self-test failure at POST. The code originates in the Host computer (HP Workstation x2100) section of the Host Computer (HP Workstation x2100 EMU/BIOS). Severity level warning (Warning — recoverable condition).

Reported action: Check that the keyboard is correctly connected; if it fails its internal self-test, replace the keyboard. If replacing the keyboard does not resolve it, change the system board, per the related MaxiLife Keyb. Ctrl. Test entry..

Documented as EMU code 0101h in the Error Message Summary table of the HP Workstation x2100 Troubleshooting Guide, part of the Philips Infinion 1.5T MR system service documentation set. Cause/Symptom: "Keyboard self-test failure"; short message: "Keyboard error". Not confirmed by this document for other Philips systems.

Full troubleshooting for 0101H

The complete diagnosis, cause tree, and step-by-step remedy for the Host computer (HP Workstation x2100) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Host computer (HP Workstation x2100) codes

Discuss 0101H with other field engineers in the FSELIB forum

Source: Philips Infinion 1.5T MR system service documentation — HP Workstation x2100 Technical Reference/Troubleshooting Guide (Chapter 4, Error Message Summary). Reference information re-authored by FSELIB — not a copy of the OEM manual.