Pular para o conteúdo principal
AlarmRTM — ELCTL for Genesis

FC040000Error in recon-process — RTM ELCTL (Genesis)

Canon Medical (Toshiba) CT error code, reported by the RTM (real time manager) — ELCTL for Genesis.

What FC040000 means

On Aquilion RX / V4-62R004 and Aquilion ONE / V4.2 CT systems, status code FC040000 indicates: Error in recon-process — RTM ELCTL (Genesis). The code originates in the RTM — ELCTL for Genesis section of the RTM (real time manager) — ELCTL for Genesis. Severity level A0 (Alarm).

Reported action: Record the fault time and identify the study whose reconstruction failed, then retry the reconstruction from the stored raw data and check the RTM error log entries logged alongside the alarm. If reconstruction fails again on the same data, follow the RTM troubleshooting procedure in the service manual..

Toshiba/Canon Aquilion RX (V4-62R004) and Aquilion ONE (V4.2) CT systems. The service error-message table lists this code under the "RTM ELCTL for Genesis" device group (set 203), with the OEM text "Error in recon-process." The same message text also appears under codes FC050000, FC060000 and FC070000 in that same device group.

Full troubleshooting for FC040000

The complete diagnosis, cause tree, and step-by-step remedy for the RTM — ELCTL for Genesis subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM — ELCTL for Genesis codes

Related service procedures

Discuss FC040000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, Aquilion RX V4-62R004 / Aquilion ONE V4.2). Reference information re-authored by FSELIB — not a copy of the OEM manual.