Notification / statusCouch / table

3341Lateral movement of couch error, COMP error of warmup

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 3341 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 3341 indicates: Lateral movement of couch error, COMP error of warmup. The code originates in the Couch / table section of the console error system (EMS). Severity level info (Notification / status).

Reported action: Lateral couch movement error detected. The couch can be used in safety mode. Drive the couch fully to the right then fully to the left to reset. If the problem persists call Service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.

Full troubleshooting for 3341

The complete diagnosis, cause tree, and step-by-step remedy for the Couch / table subsystem are documented in the FSELIB Canon Medical (Toshiba) service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related Couch / table codes

Related service procedures

Discuss 3341 with other field engineers in the FSELIB forum

Source: errmsg.eng — Toshiba/Canon CT console (EMS) error message table. Reference information re-authored by FSELIB — not a copy of the OEM manual.