Pular para o conteúdo principal
Notification / statusCouch / table

3318Vertical movement of couch error, Limit sensor error(+/-: Encoder, +: Photo sensor)

Canon Medical (Toshiba) CT error code, reported by the console error system (EMS).

What 3318 means

On Aquilion 16 and Aquilion RX and Aquilion ONE and Aquilion S16 and Asteion Super4 and Activion16 CT systems, status code 3318 indicates: Vertical movement of couch error, Limit sensor error(+/-: Encoder, +: Photo sensor). The code originates in the Couch / table section of the console error system (EMS). Severity level info (Notification / status).

Reported action: Couch error detected. The couch can be used in safety mode. Drive the couch to maximum height and lower to minimum height to reset. If the problem persists call Service..

Reported by the Toshiba/Canon CT console error system (errmsg.eng / EMS error message table), shared across the Aquilion, Asteion and Activion CT platforms.

Full troubleshooting for 3318

The complete diagnosis, cause tree, and step-by-step remedy for the Couch / table subsystem are documented in the FSELIB Canon Medical (Toshiba) service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related Couch / table codes

Related service procedures

Discuss 3318 with other field engineers in the FSELIB forum

Source: errmsg.eng — Toshiba/Canon CT console (EMS) error message table. Reference information re-authored by FSELIB — not a copy of the OEM manual.