HC11017 — No connection detected between MEC and THIB boards
Agfa CT error code, reported by the TH Module — Connectivity (TH ↔ THIB ↔ MEC).
What HC11017 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11017 indicates: No connection detected between MEC and THIB boards. The code originates in the TH Module — Connectivity (TH ↔ THIB ↔ MEC) section of the TH Module — Connectivity (TH ↔ THIB ↔ MEC). Severity level alarm (Alarm).
Reported action: Check the flat-cable connection from P38 on the MEC board to P2 on THIB — reconnect or replace the cable if not OK. If the code persists, replace the MEC board..
Raised on the Drystar 5500/5503 when no connection is detected between the MEC board and the THIB (Thermal Head Interface Board), shown on the local display as "Call service".