HC11015 — Thermal printhead data line 1 (P1 to MEC P22) connectivity fault
Agfa CT error code, reported by the TH Module — Connectivity (TH ↔ THIB ↔ MEC).
What HC11015 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11015 indicates: Thermal printhead data line 1 (P1 to MEC P22) connectivity fault. The code originates in the TH Module — Connectivity (TH ↔ THIB ↔ MEC) section of the TH Module — Connectivity (TH ↔ THIB ↔ MEC). Severity level alarm (Alarm).
Reported action: Check connectors P1 (center of the thermal printhead) and P22 (MEC board) — reconnect if not or badly connected. If the code persists, replace the data cable between P1 and P22; if it still persists after the cable swap, replace the thermal printhead itself..
Raised on the Drystar 5500/5503 when the connectivity of the data cable between P1 on the thermal printhead and P22 on the MEC board is bad, shown on the local display as "Call service".